Easy Inquiries with Wi.Tec

Gas measurements are a complex matter. For your inquiry, please use our super simple application questionnaire or send us an informal email to get in touch with us quickly.

INFRA.sens AK2/P – SF6

Gases:

Table of Contents

Features & Benefits

  • Rugged sensor design
  • No cross-sensitivity to CF4
  • Low power consumption <2W @ 24V
  • Different Interfaces (RS232, CANbus, Analog)
  • Pressure compensated (600-1200 mbar)
  • Inklusive MARS-Tool (Wi.Tec Software)

Sensor Configurations

    Gas Channel 1 (*) Gas Channel 2 (*) Gas Channel 3 (*) Gas Channel 4 (*) Options (**)
Single Gas Module 1.1 O2 P H
(*) If several gases are displayed in one column, only one gas can be selected per column
(**) O2 = Oxygen, P = Pressure, H = Humidity

List of measurement ranges

  Measuring range (*)
100
Vol.%
50
Vol.%
30
Vol.%
20
Vol.%
10
Vol.%
5
Vol.%
1
Vol.%
5,000
ppm
2,000
ppm
1,000
ppm
500
ppm
300
ppm
100
ppm
50
ppm
10
ppm
1
ppm
SF6
(*) Full scale value (F.S.)
For other measuring ranges please refer to our further datasheets

Technical Data

General features

Measurement principleNon-dispersive infrared (NDIR); dual beam;  dual wave lenghts
Measurement rangesee list of measurement ranges
Gas flow0.1 –1.5 l/min
Dimensions228.15mm x 120mm x 76mm
Weightapprox. 490g
Tube connector4/6mm tube
Lifetime of IR radiation source> 40 000h

Measuring response1

Warm-up time1 min (initial), <15 min2
Response time (t90)1.5s –15s3
Detection limit (3·σ)≤ 0,1Vol.%4
Linearity error≤ 1.0Vol.% (0.5% M.V. | 0.5% F.S.)
Repeatibility≤ ±0.2Vol.%
Zerostability< ±2% F.S./24h
Spanstability< ±2% F.S./24h
Temp. Influence zero≤ 0.5Vol.%/10K5
Temp. Influence span≤ 0.5Vol.%/10K5
Cross sensitivity≤ 0.1Vol.% @100Vol.% CF46
Pressure influence≤ 0.1%/10hPa of reading7

Electrical inputs and outputs

Supply voltage24 (15 –30) VDC
Supply current (peak)< 0.1A
Average power consumption< 2W
Digital output signalRS 232 (ASCII) or CANbus

Climatic conditions

Operating temperature5 –45 °C8
Storage temperature-20 –60 °C
Air pressure600 –1200 hPa (mbar)
Ambient humidity0 –95% rel. humidity (not condensing)
  • F.S. full scale
  • M.V measured value
  • 1 related to Pa= 1020hPa ; Ta= 25°C ; flow = 1l/min
  • 2 full specification, demands to environmental conditions
  • 3 depends on digital filter settings
  • 4 at zero point
  • 5 with zero and span temperature compensation
  • 6 other gases on request
  • 7 with pressure compensation
  • 8 stable climatic conditions recommended, please check dew point considerations