Software Download

Please leave your name and mail address. We will send you a link for the software download.

Name(Required)
Email(Required)
Privacy(Required)
This field is for validation purposes and should be left unchanged.

Easy Inquiries with Wi.Tec

Gas measurements are a complex matter. For your inquiry, please use our super simple application questionnaire or send us an informal email to get in touch with us quickly.

ULTRA.sens® AK100 H2S

Gases:

Table of Contents

Features & Benefits

  • High dynamic range
  • Fast response time
  • Applicable for continuous measurements
  • Different Interfaces (RS232, CANbus)

Sensor Configurations

    Gas Channel 1 (*) Gas Channel 2 (*) Gas Channel 3 (*) Gas Channel 4 (*) Options (**)
Single Gas Module 1.1 O2 P H
Dual Gas Module 2.1 O2 P H
2.2 O2 P H
(*) If several gases are displayed in one column, only one gas can be selected per column
(**) O2 = Oxygen, P = Pressure, H = Humidity

List of Measurement Ranges

  Measuring range (*)
100
Vol.%
50
Vol.%
30
Vol.%
20
Vol.%
10
Vol.%
5
Vol.%
1
Vol.%
5,000
ppm
2,000
ppm
1,000
ppm
500
ppm
300
ppm
100
ppm
50
ppm
10
ppm
1
ppm
H2S
SO2
(*) Full scale value (F.S.)
For other measuring ranges please refer to our further datasheets

Technical Data

General features
Measurement principle Non-dispersive ultraviolett (NDUV); dual beam
Measurement range see list of measurement ranges
Gas flow 0.1 – 1.5 l/min
Dimensions 300mm x 100mm x 81mm
Weight approx. 1670g
Tube connector 4/6mm tube
Lifetime of UV radiation source > 8 000h1
Measuring response
Warm-up time 1 min (initial), <15 min 3
Response time (t90) 1.5s – 15s4
Detection limit (3·σ) < 1ppm5
Linearity error < ± 1% F.S.
Repeatibility ± 0.5% F.S.
Long term stability (zero) < ± 1% F.S./24h
Long term stability (span) < ± 1% F.S./month
Temp. Influence zero < 1% F.S./10K
Temp. Influence span < 2% F.S./10K6
Cross sensitivity < 2% F.S. 7
Pressure influence < 1.5%/10hPa of reading8
related to Pa = 1020hPa ; Ta= 25°C ; flow = 1l/min // 1 F.S. full scale // 2 full specification, demands to environmental conditions // 3 depends on digital filter settings // 4 at zero point // 5 with span temperature compensation // 6 to each calibrated gas channel, other gases on request // 7 without pressure compensation
Electrical inputs and outputs
Supply voltage 24 (15 – 30) VDC
Supply current (peak) < 0.4A
Average power consumption < 7.5W
Digital output signal RS 232 (ASCII) or CAN bus
Climatic conditions
Operating temperature 5 – 45 °C9
Storage temperature -20 – 60 °C
Air pressure 800 – 1200 hPa (mbar)