Non-dispersive infrared (NDIR); dual beam; dual to quad wavelenghts
Measurement range
see list of measurement ranges
Gas flow
0.1 – 1.5 l/min
Dimensions
225mm x 120mm x 74.4mm
Weight
approx. 520g
Tube connector
4/6mm tube
Lifetime of IR radiation source
> 40 000h
Measuring response*
Warm-up time
1 min (initial), <15 min 2
Response time (t90)
1.5s – 15s 3
Detection limit (3·σ)
< 0,5 %F.S. 4
Linearity error
< ± 1 %F.S.
Repeatibility
± 0.5 %F.S.
Long term stability (zero)
< ± 2 %F.S. /week
Long term stability (span)
< ± 2 %F.S. /month
Temp. Influence zero
< 1 %F.S. / 10K
Temp. Influence span
< 1 %F.S. / 10K 5
Cross sensitivity
< 2 %F.S. 6
Pressure influence
< 1.5% / 10hPa of reading
related to Pa = 1020hPa ; Ta= 25°C ; flow = 1l/min // 1 F.S. full scale // 2 full specification, demands to environmental conditions // 3 depends on digital filter settings // 4 at zero point // 5 with span temperature compensation // 6 to each calibrated gas channel, other gases on request // 7 without pressure compensation