ULTRA.sens® AK100 TBH

Gases:

Table of Contents

Features & Benefits

  • Simultaneous NO₂ and SO₂ analysis
  • Temperature controlled to 50°C
  • Easy integration via mounting bracket
  • Fast response time
  • No influence of gas humidity

Sensor Configurations

    Gas Channel 1 (*) Gas Channel 2 (*) Gas Channel 3 (*) Gas Channel 4 (*) Options (**)
Single Gas Module 1.1 O₂ P H
Dual Gas Module 2.1 O₂ P H
Triple Gas Module 3.1 O₂ P H
(*) If several gases are displayed in one column, only one gas can be selected per column
(**) O₂ = Oxygen, P = Pressure, H = Humidity

List of Measurement Ranges

  Measuring range (*)
100
Vol.%
50
Vol.%
30
Vol.%
20
Vol.%
10
Vol.%
5
Vol.%
1
Vol.%
5,000
 ppm
2,000
 ppm
1,000
 ppm
500
 ppm
300
 ppm
100
 ppm
50
 ppm
10
 ppm
1
 ppm
NO₂
SO₂
NO
(*) Full scale value (F.S.)
For other measuring ranges please refer to our further datasheets

Technical Drawing (Dimensions)

TBH V5 H Dimensions

Technical Data

General features*
Measurement principle Ultraviolet resonance absorption spectroscopy (UVRAS) /
Non-dispersive ultraviolet (NDUV) ; dual beam
Measurement range see list of measurement ranges
Gas flow 0.1 – 1.5l/min
Dimensions 345mm × 153mm × 103mm
Weight 3270g
Tube connector 4/6mm tube
Lifetime of UV radiation source >20 000h [LED; NO2 / SO2]
>8 000h1 [EDL; NO]
Measuring response*
Warm-up time ≤ 45min (initial)3
Response time (t90) 1.5s – 15s4
Detection limit (3·σ) ≤ 1 ppm5
Linearity error ≤ ± 1 %F.S.
Repeatibility ± 0.5 %F.S.
Long term stability (zero) < 3 ppm / 24h3
Long term stability (span) <± 1 %F.S. /month
Temp. Influence zero < 0.5 %F.S. /10K
Temp. Influence span < 0.5 %F.S. /10K
Cross sensitivity 100 ppm SO2 < 2 ppm
500 ppm NO2 < 2 ppm
20°C D.P. H2O < 10 ppm
100 ppm N2O < 10 ppm6
Pressure influence <1.5%/10hPa of reading 7 of reading
related to Pa = 1020hPa ; Ta= 25°C ; flow = 1l/min // 1 F.S. full scale // 2 full specification, demands to environmental conditions // 3 depends on digital filter settings // 4 at zero point // 5 with span temperature compensation // 6 to each calibrated gas channel, other gases on request // 7 without pressure compensation
Electrical inputs and outputs
Supply voltage 24 (20 – 30) VDC
Supply current (peak) 1,5 A
Inrush current 0.2 – 0.7 A
Power consumption (average) 36 W
Digital output signal RS 232 (ASCII) or CAN bus
Climatic conditions
Operating temperature* 5 – 40 °C
Storage temperature -20 – 60 °C
Air pressure 600 – 1200 hPa (mbar)
Ambient humidity 0 – 95 % rel. humidity (not condensing)
*stable climatic conditions recommended, please check dewpoint considerations