An EDL (electrodeless gas discharge lamp) is used to detect NO and H₂S. A combination of both technologies (UVRAS & UVLEDs) allows simultaneous gas analysis of NO, NO₂ and SO₂ in the low ppm range, which is particularly important in flue gas analysis (CEM).
In the EDL, N₂ and O₂ are converted to NO and generate selective UV radiation. This radiation enables a NO measurement that is free of cross-sensitivities. This method is referred to as UV resonance absorption spectroscopy (UVRAS).
Gas Channel 1 (*) | Gas Channel 2 (*) | Gas Channel 3 (*) | Gas Channel 4 (*) | Options (**) | ||||
---|---|---|---|---|---|---|---|---|
Single Gas Module | 1.1 | | P | H | ||||
Dual Gas Module | 2.1 | | | P | H | |||
Triple Gas Module | 3.1 | | | | P | H |
Measuring range (*) | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
100 Vol.% | 50 Vol.% | 30 Vol.% | 20 Vol.% | 10 Vol.% | 5 Vol.% | 1 Vol.% | 5,000 ppm | 2,000 ppm | 1,000 ppm | 500 ppm | 300 ppm | 100 ppm | 50 ppm | 10 ppm | 1 ppm | |
NO | ||||||||||||||||
NO₂ | ||||||||||||||||
SO₂ |
General features* | |
---|---|
Measurement principle | Ultraviolet resonance absorption spectroscopy (UVRAS) / Non-dispersive ultraviolet (NDUV) ; dual beam |
Measurement range | see list of measurement ranges |
Gas flow | 0.1 – 1.5l/min |
Dimensions | 345mm × 153mm × 103mm |
Weight | 3270g |
Tube connector | 4/6mm tube |
Lifetime of UV radiation source | >20 000h [LED; NO₂ / SO₂] > 8 000h1 [EDL; NO] |
Measuring response* | |
---|---|
Warm-up time | ≤ 45min (initial)3 |
Response time (t90) | 1.5s – 15s4 |
Detection limit (3·σ) | ≤ 1 ppm5 |
Linearity error | ≤ ± 1 %F.S. |
Repeatibility | ± 0.5 %F.S. |
Long term stability (zero) | < 3 ppm / 24h3 |
Long term stability (span) | <± 1 %F.S. /month |
Temp. Influence zero | < 0.5 %F.S. /10K |
Temp. Influence span | < 0.5 %F.S. /10K |
Cross sensitivity | 100 ppm SO₂ < 2 ppm 500 ppm NO₂ < 2 ppm 20°C D.P. H₂O < 10 ppm 100 ppm N₂O < 10 ppm6 |
Pressure influence | <1.5%/10hPa of reading 7 of reading |
Electrical inputs and outputs | |
---|---|
Supply voltage | 24 (20 – 30) VDC |
Supply current (peak) | 1,5 A |
Inrush current | 0.2 – 0.7 A |
Power consumption (average) | 36 W |
Digital output signal | RS 232 (ASCII) or CAN bus |
Climatic conditions | |
---|---|
Operating temperature* | 5 – 40 °C |
Storage temperature | -20 – 60 °C |
Air pressure | 600 – 1200 hPa (mbar) |
Ambient humidity | 0 – 95 % rel. humidity (not condensing) |